Cod. TCFEMD342 - H2
Cod. TCFEMD344 - H4
Cod. TCFEMD346 - H6
Cod. TCFEMA15°342 - H2
Cod. TCFEMA15°344 - H4
Cod. TCFEMA15°346 - H6
Cod. TCFEMA25°342 - H2
Cod. TCFEMA25°344 - H4
Cod. TCFEMA25°346 - H6
Cod. TCFECA34N
Cod. TCFECA34R
Cod. TCFEMP34N
Cod. TCFEMP34R
Cod. TCFEUC34N
Cod. TCFEMS341 - H1
Cod. TCFEMS342 - H2
Cod. TCFEMS343 - H3
Cod. TCFEMS344 - H4
Cod. MSCAP01
Cod. MSCON01
Cod. TCFE00°3420
Cod. TCFE17°3420
Cod. TIS900002
Cod. TIS100222
Cod. TIS100003
Cod. TPOMFA
Cod. TCFETR34
Cod. TCFETRST34
Cod. TCFEVTRST
Code TCFEAN34
Cod. TCFEVG342 - H2
Cod. TCFEVG344 - H4
Cod. TCFEVG346 - H6
Cod. TIS900003
Cod. TIS900004
Cod. TIS900001
Cod. TIS100304
Cod. TCFEVPL
Cod. TCFEVPR
Cod. TPOMFA
Cod. TIS100001
Cod. TIS900002
Cod. TPRFR
Cod. TFRLANC
Cod. TFRC20
Cod. TFRC28
Cod. TCEIMS32
Cod. TDR12C
Cod. TDR12L
Cod. TCR12L
TCONTR
Cod. TIND
Cod. TMPRF
Cod. TCHBM
Cod. TDINA
Cod. TCRMFA
Cod. 771CEF
Cod. TSTOPC20XX
(07/85/10/11/13/15)
Cod. TSTOPC28XX
(07/85/10/11/13/15)
For the correct insertion of the cylindrical external hexagon Ø 3.25 implant, we recommend following the surgical protocol:
1. Use the lanceolate drill (TFRLANC) up to the marking (recommended n. revs = 800 rpm).
2. Use the 2 mm start-up drill (TFRC20) with the appropriate stop of the same height as the implant being inserted (recommended n. revs = 800 rpm).
3. Insert the parallelism indicator and check for any disparallelism.
4. Use the final cylindrical drill Ø 2.8 (TFRC28) with the appropriate stop (recommended n. revs = 200-400 rpm).
5. In the case of D1 bone density, use the threadformer (TCEIMS32)
6. Insert the implant. Recommended n. of revs 15-25 rpm.
7. Close with the screw cap included in the package.